Gentec-EO HP User Manual
Page 24

Version 2.2
20
HP Series Instruction Manual
Gentec Electro-Optics Inc. All rights reserved
Laser
HP60A-10KW-GD (< Ø 35 mm)
HP100A-12KW-HD
HP60A-10KW-GD (> Ø 35 mm)
Beam
Damage
Damage 1/e²
Min. 1/e²
Damage
Damage 1/e²
Min. 1/e²
Power
Threshold
6
Beam Diam.
7
,
8
Beam Diam.
Threshold
Beam Diam.
Beam Diam.
[kW]
[kW/cm²]
[cm]
[cm]
[kW/cm²]
[cm]
[cm]
0.5
40
0.2
0.3
16
0.3
0.4
1
35
0.3
0.4
14
0.4
0.6
2
30
0.4
0.6
12
0.7
0.9
3
25
0.6
0.8
9.7
0.9
1.3
4
20
0.7
1.0
7.9
1.1
1.6
5
17
0.9
1.2
6.5
1.4
2.0
6
15
1.0
1.4
5.6
1.7
2.3
7
13
1.2
1.7
4.9
1.9
2.7
8
12
1.3
1.8
4.3
2.2
3.1
9
11
1.4
2.0
4.0
2.4
3.4
10
10
1.6
2.3
3.5
2.7
3.8
11
3.0
3.1
4.3
12
2.5
3.5
4.9
In the event of major damage to the coating, the HP Series sensors can be recoated. Contact your
local Gentec-EO representative for information on repair and recalibration. See p. ii Contacting
Gentec Electro-Optics Inc.
6 Peak Intensity.
7 For Gaussian beam profile, the peak intensity is twice the beam power.
8 Diameter of a circle corresponding to 86% of the entire beam power.
9 Including a security factor of 50%