Gentec-EO HP User Manual

Page 24

Advertising
background image

Version 2.2

20

HP Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

Laser

HP60A-10KW-GD (< Ø 35 mm)

HP100A-12KW-HD

HP60A-10KW-GD (> Ø 35 mm)

Beam

Damage

Damage 1/e²

Min. 1/e²

Damage

Damage 1/e²

Min. 1/e²

Power

Threshold

6

Beam Diam.

7

,

8

Beam Diam.

7

,

8

,9

Threshold

6

Beam Diam.

7

,

8

Beam Diam.

7

,

8

,

9

[kW]

[kW/cm²]

[cm]

[cm]

[kW/cm²]

[cm]

[cm]

0.5

40

0.2

0.3

16

0.3

0.4

1

35

0.3

0.4

14

0.4

0.6

2

30

0.4

0.6

12

0.7

0.9

3

25

0.6

0.8

9.7

0.9

1.3

4

20

0.7

1.0

7.9

1.1

1.6

5

17

0.9

1.2

6.5

1.4

2.0

6

15

1.0

1.4

5.6

1.7

2.3

7

13

1.2

1.7

4.9

1.9

2.7

8

12

1.3

1.8

4.3

2.2

3.1

9

11

1.4

2.0

4.0

2.4

3.4

10

10

1.6

2.3

3.5

2.7

3.8

11

3.0

3.1

4.3

12

2.5

3.5

4.9


In the event of major damage to the coating, the HP Series sensors can be recoated. Contact your
local Gentec-EO representative for information on repair and recalibration. See p. ii Contacting
Gentec Electro-Optics Inc.

6 Peak Intensity.
7 For Gaussian beam profile, the peak intensity is twice the beam power.
8 Diameter of a circle corresponding to 86% of the entire beam power.
9 Including a security factor of 50%