Gentec-EO HP User Manual

Page 9

Advertising
background image

Version 2.2

5

HP Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

HP…

HP60A-10KW-GD

HP100A-12KW-HD

Aperture Diameter

60 mm conical

Optimized for 35 mm

100 mm

Spectral Range

0.8 - 12 µm

0.19 - 20 µm

Power Noise Level

± 10 W

± 10 W

Typical Rise Time

(0

– 95 %)

11 s

9 s

Sensitivity

0.2 mV/W

0.15 mV/W

Calibration Uncertainty

± 5 % @ 1.064 µm

Linearity

with Power

± 2%

Repeatability

(Precision)

± 2%

Min. Average Power

(continuous)

300 W

7

Max. Average Power

(continuous)

10 000 W

12 000 W

Max. Average Power

(5 min.)

10 000 W

12 000 W

Max. Average Power

Density

1 2

< Ø 35 mm: 10 kW/cm² (10 kW)

> Ø 35 mm: 3.5 kW/cm² (10 kW)

16 kW/cm² (0.5 kW)

6.5 kW/cm² (5 kW)

3.5 kW/cm² (10 kW)

Dimension

(mm)

127 (H) x 127 (W) x 90 (D)

127 (H) x 127 (W) x 70 (D)

Weight

(head only)

≈5 kg

3.3 kg

Cooling

Water

Required cooling flow

6

– 10 LPM

< ± 1 LPM/min

3

Temp of cooling water

15 - 25 °C

< ± 3 °C/min

3

Pressure Drop

< 11 psi

Recommended load

Impedance

> 100 kΩ

Output Impedance

< 675

Ω

Linearity

vs beam dimension

4

< Ø 35 mm: ± 0.5 %
> Ø 35 mm: ± 1.5 %

± 1.0 %

Linearity

vs beam position

± 3.0 %

5

± 1.5 %

6

± 3.0 %

5

PCB electrical supply

USB or

Gentec-EO monitors

Maximum current

consumption

30 mA

Max output signal

2.0 V

1

: Refer to section 4 for more details on the appropriate beam size in function of measured power.

2

: 1.064, 1.07-1.08 & 10.6 µm.

3

: On a period > 1 minute.

4

: For a centered beam size from 10% to 80% of the aperture area.

5

: For a beam size of 20% of the aperture area, moved across 80% of the aperture area.

6

: For a beam size of 10% of the aperture area, moved across 40% of the aperture area.

7

: For lower power, consult Gentec-eo.

Specifications subject to change without notice.

Advertising