Gentec-EO HP User Manual
Page 9

Version 2.2
5
HP Series Instruction Manual
Gentec Electro-Optics Inc. All rights reserved
HP…
HP60A-10KW-GD
HP100A-12KW-HD
Aperture Diameter
60 mm conical
Optimized for 35 mm
100 mm
Spectral Range
0.8 - 12 µm
0.19 - 20 µm
Power Noise Level
± 10 W
± 10 W
Typical Rise Time
(0
– 95 %)
11 s
9 s
Sensitivity
0.2 mV/W
0.15 mV/W
Calibration Uncertainty
± 5 % @ 1.064 µm
Linearity
with Power
± 2%
Repeatability
(Precision)
± 2%
Min. Average Power
(continuous)
300 W
7
Max. Average Power
(continuous)
10 000 W
12 000 W
Max. Average Power
(5 min.)
10 000 W
12 000 W
Max. Average Power
Density
1 2
< Ø 35 mm: 10 kW/cm² (10 kW)
> Ø 35 mm: 3.5 kW/cm² (10 kW)
16 kW/cm² (0.5 kW)
6.5 kW/cm² (5 kW)
3.5 kW/cm² (10 kW)
Dimension
(mm)
127 (H) x 127 (W) x 90 (D)
127 (H) x 127 (W) x 70 (D)
Weight
(head only)
≈5 kg
3.3 kg
Cooling
Water
Required cooling flow
6
– 10 LPM
< ± 1 LPM/min
3
Temp of cooling water
15 - 25 °C
< ± 3 °C/min
3
Pressure Drop
< 11 psi
Recommended load
Impedance
> 100 kΩ
Output Impedance
< 675
Ω
Linearity
vs beam dimension
4
< Ø 35 mm: ± 0.5 %
> Ø 35 mm: ± 1.5 %
± 1.0 %
Linearity
vs beam position
± 3.0 %
5
± 1.5 %
6
± 3.0 %
5
PCB electrical supply
USB or
Gentec-EO monitors
Maximum current
consumption
30 mA
Max output signal
2.0 V
1
: Refer to section 4 for more details on the appropriate beam size in function of measured power.
2
: 1.064, 1.07-1.08 & 10.6 µm.
3
: On a period > 1 minute.
4
: For a centered beam size from 10% to 80% of the aperture area.
5
: For a beam size of 20% of the aperture area, moved across 80% of the aperture area.
6
: For a beam size of 10% of the aperture area, moved across 40% of the aperture area.
7
: For lower power, consult Gentec-eo.
Specifications subject to change without notice.