Operating procedure – Robinair 16900 Refrigerant Identifier User Manual

Page 14

Advertising
background image

12

7. The results of the analysis are now presented to the user on the

instrument display.

PASS: Displayed when a sample is detected to have 98% or

greater of R-12 or R-134a by weight. The refrigerant type and air

concentration are also displayed.

FAIL: Displayed when a sample is detected to be a blend of

R-12 and R-134a refrigerants with neither R-12 or R-134a having

98% or greater concentration. The weight percentages of R-12,

R-134a, and air are displayed.

FAIL CONTAMINATED: Displayed when unknown refrigerant

types, such as R-22 or hydrocarbons, have been detected

in concentrations of 4% or greater by weight. No weight

concentrations of refrigerants or air are provided in this mode.

NO REFRIGERANT—CHK HOSE CONN: Displayed when a

sample is detected containing 90% or greater air by weight. This

condition usually occurs when the R-134a sample hose coupler has

not been opened, the sample hose has not been connected to the

sample source, or if there is no refrigerant present in the sample

source.

8. The instrument keeps the analysis results on the display until the

user presses the

A button. Once the A button is pressed, follow the

directions on the screen.

9. If another sample run is desired, follow instructions from Step 5.
If another sample run is not desired, disconnect the power harness

from the vehicle or wall outlet to the end the routine.