Error codes, Warning – Robinair 16900 Refrigerant Identifier User Manual

Page 18

Advertising
background image

16

AIR SENSOR FAULT: Signifies an air detection sensor calibration fault.
Corrective Action:
1. Verify the air intake port and sample exhaust port of the instrument

are not clogged or obstructed.

2. If the instrument is being used in an enclosed area, oxygen-

depleting compounds, such as carbon dioxide or carbon monoxide,

may accumulate in the surrounding atmosphere and be drawn into

the intake port during calibration. Move the instrument to a location

with fresh ambient air, free of oxygen-depleting compounds.

Use the instrument in locations that provide adequate ventilation

to prevent the accumulation of oxygen-depleting compounds in the

surrounding atmosphere.

WARNING

If error codes indicate that oxygen-depleting compounds are in

the surrounding atmosphere, take immediate action to rid the area

of these compounds, and provide adequate ventilation to prevent

such accumulations. The accumulation of oxygen-depleting

compounds in the surrounding atmosphere present a hazard to

human health.

3. To clear the error code, power down the instrument. The error code

automatically clears.

RECALIBRATION REQUIRED: The prior calibration is too old and

requires refreshing. This message appears whenever a 3-minute period

elapses from the prior calibration.
Corrective Action:
1. Remove the sample hose connection from the refrigerant source if

so connected. Depress the

A button to initiate a new calibration.