Error codes – Robinair 16900 Refrigerant Identifier User Manual

Page 17

Advertising
background image

16900 Refrigerant Identifier

15

DETECTOR FAULT: Signifies internal infrared sensor failure due to

temperature extremes.
Corrective Actions:
1. Allow the instrument to stabilize at room temperature for at least

30 minutes before attempting use.

2. To clear the error code, power down the instrument. The error code

automatically clears.

CALIBRATION FAULT: Signifies an infrared calibration fault.
Corrective Actions:
1. Verify the air intake port and sample exhaust port of the instrument

are not clogged or obstructed.

2. If the instrument is being used in an enclosed area, refrigerant

vapors may build up in the surrounding atmosphere and be drawn

into the intake port during calibration. Move the instrument to a

location with fresh ambient air, free of refrigerant vapors.

Use the instrument in locations that provide adequate ventilation to

prevent the accumulation of refrigerant vapors in the surrounding

atmosphere.

3. Hang the unit so there is a flow of fresh air around it.
4. To clear the error code, power down the instrument. The error code

automatically clears.

Error Codes