Error codes, Error codes and corrective actions – Robinair 16900 Refrigerant Identifier User Manual

Page 16

Advertising
background image

14

ERROR CODES AND CORRECTIVE ACTIONS

The instrument is supplied with self-diagnostic software to aid in

determining problems. The software will supply error messages that

direct the user toward specific corrective actions in the event of system

failure. Should corrective measures not correct the problem, the user is

advised to contact Robinair at 1-800-822-5561 for assistance.
SAMPLE FAULT: Signifies unstable infrared readings due to

inconsistent sample gas supply or interference from strong electrical

fields.
Corrective Actions:
1. If the error code occurs during a sample cycle, verify the sample

hose is receiving sample gas from the source connection of at least

5 psig. Verify the instrument sample exhaust port and air intake

port are not clogged or restricted.

2. If the error code occurs during a calibration cycle, verify the air

intake port and sample exhaust port of the instrument are not

clogged or obstructed.

3. Keep the unit away from sources of strong electrical fields, such as

large compressors, running vehicles, etc. Move the instrument 3–5

feet away from the electrical field source, and try again.

4. If the instrument is being used outside of the operating temperature

range of 39–122°F (4–50°C), allow the instrument to warm or cool

to the operating temperature range, and try again. The instrument

may require additional warm-up time after “power up” before use.

5. To clear the error code, power down the instrument. The error code

automatically clears.