Manual-zero – SKC Limited EPAM-5000 User Manual
Page 33
                                    Advertising
                                    
                                        
                                    
                                
                                 
Manual-Zero,
Continued
3
Select Special Functions from the Main Menu.
4
Select System Options.
5
Select Extended Options.
6
Select Calibration. 
 
Result: Screen appears with manual zero at the top. 
7
Select Manual-Zero. 
Select Manual-Zero again. 
 
Note: Wait approximately 99 seconds. The unit automatically 
executes the steps necessary to reestablish the baseline. 
 
Result: The Main Menu is displayed when the manual-zero 
process is complete. 
 
Notes: 
The EPAM-5000 has an auto zero purging feature that 
automatically purges the sensor and performs an auto zero to 
reestablish the baseline approximately every 30 minutes. An X 
inside a box appears in right hand corner of the display when the 
instrument is auto zeroing. 
 
 
3 - 8