Probe connection to pcf200, Figure 12. pcf200 deskew fixture, Qphy-ddr3 software option – Teledyne LeCroy QPHY-DDR3 User Manual

Page 9

Advertising
background image

QPHY-DDR3 Software Option

917717 Rev C

9

Figure 1. PCF200 Deskew Fixture

A SMA male to BNC male 50-ohm cable is required to perform the calibration.

System assembly is accomplished in the following steps:

1.

Connect the BNC end of the 50 ohm cable to the oscilloscope AUX IN input.

2.

Connect the SMA end of the 50 ohm cable to the SMA female connector on the PCF200 fixture.

On oscilloscope models with a dedicated Fast Edge SMA output,

3a.

Connect the PCF200 SMA male connector to the oscilloscope Fast Edge SMA output.

Skip to the next section.

On oscilloscope models without a dedicated Fast Edge SMA output,

3b.

Using an BNC to female SMA adapter, connect the PCF200 SMA male connect to the
oscilloscope AUX OUT output.

4.

Under Utilities

Utilities Setup, select Fast Edge from within the Aux Output tab.

Probe Connection to PCF200

The PCF200 provides multiple probe connectors for various kinds of probes. There are 2 circuits
depending on the type of probes to calibrate:

• The upper circuit is for Solder-In (SI) and Quick-Connect (QC) probes. This circuit can also be

used for AT probes using the designated area to apply the probe tips.

• The lower circuit is for Square-Pin (SP) probes.

Probes are connected electrically in a single-ended arrangement: the positive (+) side of the probe must
be connected to the signal trace, while the negative (-) side is connected to the ground plane. The
positive polarity marking is located on the tip end of the probe in white as a plus sign.

Advertising