Command Reference

WaveStation 3000 Arbitrary Waveform Values

Index

Name

Index

Name

Index

Name

Index

Name

Index

Name

0

StairUp

8

Sqrt

16

Gmonopuls 24

Blackman

32

Asin

1

StairDn

9

X^2

17

Cardiac

25

GaussiWin

33

Acos

2

StairUD

10

Sinc

18

Quake

26

Harris

34

Atan

3

Trapezia

11

Gaussian

19

TwoTone

27

Bartlett

35

ACot

4

ExpFall

12

Dlorentz

20

SNR

28

Tan

5

ExpRise

13

Haversine

21

Hamming

29

Cot

6

LogFall

14

Lorentz

22

Hanning

30

Sec

7

LogRise

15

Gauspuls

23

Kaiser

31

Csc

Indices 36-59 are reserved for user-defined waveforms up to 16K; indices 60-67 are for user-defined
waveforms up to 512K.

BSWV - Basic Wave Command

DESCRIPTION

Sets or retrieves basic waveform parameters.

SYNTAX

<channel>:BaSicWaVe <parameter>

<channel>:=

{C1|C2}

<parameter>:= {a parameter from the table below}

IMPORTANT: If the command doesn’t specify wave type (WVTP), the remaining parameters are
applied to the waveform type already selected on your WaveStation.

Parameters

Value

Description

WVTP

<wave type>

Where <wave type>:= {SINE, SQUARE, RAMP, PULSE, NOISE, ARB ,DC}

FRQ

<frequency>

Where <frequency>:= {default unit is Hz; minimum value 1xe-6 Hz, maximum

value depends on the model}. If WVTP is NOISE, parameter is invalid.

AMP

<amplifier>

Where <amplifier>:= {default unit is V;

WS2000 Channel 1 min. 0.004V, max. 6V; Channel 2 min. 0.004V, max. 20V

WS3000 Channel 1 and 2 min. 2mVpp, max. 20Vpp}.

If WVTP is NOISE, parameter is invalid.

OFST

<offset>

Where <offset>:= {default unit is V; maximum value depends on the

WaveStation model}. If WVTP is NOISE, parameter is invalid.

SYM

<symmetry>

Where <symmetry> :={0 to 100}. Parameter valid only when WVTP is RAMP.

DUTY

<duty cycle>

Where <duty cycle>:= {if WVTP is SQUARE, 20% to 80%; if WVTP is PULSE,

0.1% to 99.9%}.

PHSE

<phase>

Where <phase>:= {0 to 360}. If WVTP is NOISE, parameter is invalid.

STDEV

<standard

deviation>

Where <standard deviation>:= {default unit is V; min. 0.5mV; WS2000 max.

1.625V, WS3000 max. 1.599V}. Parameter is valid only when WVTP is NOISE.