Teledyne LeCroy WaveStation SCPI Command Reference Manual User Manual

Page 18

Advertising
background image

WaveStation SCPI

Parameters

Value

Description

MEAN

<mean>

Parameter is valid only when WVTP is NOISE. Where <mean>:= {default unit is

V; range depends on <standard deviation>}.

WIDTH

<width>

Value of width. Parameter is valid only when WVTP is PULSE.

RISE

<rise>

Value of rise. Parameter is valid only when WVTP is PULSE.

FALL

<fall>

Value of fall. Parameter is valid only when WVTP is PULSE.

DLY

<delay>

Where <delay>:= {default unit is S; minimum 0, maximum is Pulse Period}.

Parameter is valid only when WVTP is PULSE.

QUERY SYNTAX

<channel>:BaSicWaVe?

<channel>:=

{C1|C2}

RESPONSE

<channel>:BSWV <type>,<frequency>,<offset>,<amplifier>,<duty>,

<symmetry>,<phase>

EXAMPLE 1

Set Channel 1 basic wave type to Ramp.

C1:BSWV WVTP,RAMP

EXAMPLE 2

Set Channel 1 signal frequency to 2000Hz.

C1:BSWV FRQ,2000HZ

EXAMPLE 3

Retrieve Channel 1 Basic Wave settings.

C1:BSWV?

Return:

C1:BSWV TYPE,SINE,FRQ,1000,AMP,3,OFST,3

RELATED

ARWV, BTWV, CFG, CPL, MDWV, SWWV

16

922829 Rev C