Teledyne LeCroy WaveStation SCPI Command Reference Manual User Manual

Page 20

Advertising
background image

WaveStation SCPI

Parameters

Value

Description

(CARR,)AMP

<amplifier>

Where carrier wave <amplifier>:= {default unit is V;

WS2000 Channel 1 min. 0.004V max. 6V, Channel 2 min. 0.004V max. 20V

WS3000 Channel 1 and 2 min. 2mVpp max. 20Vpp}.

(CARR,)OFST

<offset>

Where carrier wave <offset>:= {default unit is V; maximum value depends on

the WaveStation model}.

(CARR,)SYM

<symmetry>

Where carrier wave <symmetry>:= {0% to 100%}.

(CARR,)DUTY

<duty cycle>

Where carrier wave <duty cycle>:= {if CARR,WVTP is SQUARE, 20% to

80%; if CARR,WVTP is PULSE, 0.1% to 99.9%}.

(CARR,)PHSE

<phase>

Where carrier wave <phase>:= {0 to 360}.

(CARR,)RISE

<rise>

Value of carrier wave rising edge. Parameter valid only when CARR,WVTP

is PULSE.

(CARR,)FALL

<fall>

Value of carrier wave falling edge. Parameter valid only when CARR,WVTP

is PULSE.

(CARR,)DLY

<delay>

Where carrier wave <delay>:= {default unit is S; minimum 0, maximum is

value of Pulse Period}. Parameter is valid only when CARR,WVTP is

PULSE.

(CARR,)STDEV

<standard

deviation>

Parameter is valid only when CARR,WVTP is NOISE.

(CARR,)MEAN

<mean>

Parameter is valid only when CARR,WVTP is NOISE.

QUERY SYNTAX

<channel>:BursTWaVe?

RESPONSE

<channel>:BursTWaVe <type>,<state>,<period>

EXAMPLE 1

Set Channel 1 burst waveform period to one second.

C1:BTWV PRD,1S

EXAMPLE 2

Read Channel 2 burst waveform parameters (and STATE is ON).

C2:BTWV?

Return:

C2:BURSTWAVE STATE,ON,PRD,0.01S,STPS,0,TRIG,IN,TRMD,OFF,

TIME,1,DLAY,2.4e-07S,GATE_NCYC,NCYC,PLRT,NEG,CARR,WVTP,SINE,

FRQ,1000HZ,AMP,4V,OFST,0V,PHSE,0

EXAMPLE 3

Read Channel 2 burst wave parameters (and STATE is OFF).

C2:BTWV?

Return:

C2:BURSTWAVE STATE,OFF

18