Teledyne LeCroy WaveStation SCPI Command Reference Manual User Manual
Page 20

WaveStation SCPI
Parameters
Value
Description
(CARR,)AMP
<amplifier>
Where carrier wave <amplifier>:= {default unit is V;
WS2000 Channel 1 min. 0.004V max. 6V, Channel 2 min. 0.004V max. 20V
WS3000 Channel 1 and 2 min. 2mVpp max. 20Vpp}.
(CARR,)OFST
<offset>
Where carrier wave <offset>:= {default unit is V; maximum value depends on
the WaveStation model}.
(CARR,)SYM
<symmetry>
Where carrier wave <symmetry>:= {0% to 100%}.
(CARR,)DUTY
<duty cycle>
Where carrier wave <duty cycle>:= {if CARR,WVTP is SQUARE, 20% to
80%; if CARR,WVTP is PULSE, 0.1% to 99.9%}.
(CARR,)PHSE
<phase>
Where carrier wave <phase>:= {0 to 360}.
(CARR,)RISE
<rise>
Value of carrier wave rising edge. Parameter valid only when CARR,WVTP
is PULSE.
(CARR,)FALL
<fall>
Value of carrier wave falling edge. Parameter valid only when CARR,WVTP
is PULSE.
(CARR,)DLY
<delay>
Where carrier wave <delay>:= {default unit is S; minimum 0, maximum is
value of Pulse Period}. Parameter is valid only when CARR,WVTP is
PULSE.
(CARR,)STDEV
<standard
deviation>
Parameter is valid only when CARR,WVTP is NOISE.
(CARR,)MEAN
<mean>
Parameter is valid only when CARR,WVTP is NOISE.
QUERY SYNTAX
<channel>:BursTWaVe?
RESPONSE
<channel>:BursTWaVe <type>,<state>,<period>
EXAMPLE 1
Set Channel 1 burst waveform period to one second.
C1:BTWV PRD,1S
EXAMPLE 2
Read Channel 2 burst waveform parameters (and STATE is ON).
C2:BTWV?
Return:
C2:BURSTWAVE STATE,ON,PRD,0.01S,STPS,0,TRIG,IN,TRMD,OFF,
TIME,1,DLAY,2.4e-07S,GATE_NCYC,NCYC,PLRT,NEG,CARR,WVTP,SINE,
FRQ,1000HZ,AMP,4V,OFST,0V,PHSE,0
EXAMPLE 3
Read Channel 2 burst wave parameters (and STATE is OFF).
C2:BTWV?
Return:
C2:BURSTWAVE STATE,OFF
18
922829 Rev C