BTWV - Burst Wave Command
DESCRIPTION
Sets waveform type to burst or retrieves Burst Wave settings.
SYNTAX
<channel>:BursTWaVe STATE,ON,<parameter>,<value>,…
<channel>:=
{C1|C2}
<parameter>:= {a parameter from the table below}
<value>:=
{a value from the table below}
IMPORTANT: Set STATE ON before setting other parameters. Specify carrier wave type before listing
other CARR parameters (e.g., C1:BTWV STATE,ON,CARR,WVTP,<value>....). Once set, you do not need
to repeat CARR; it is shown here to indicate which parameters can follow it.
Parameters
Value
Description
STATE
{ON|OFF}
To set or read other Burst Wave parameters, first set STATE ON.
PRD
<period>
Where <period>:= {default unit is S; WS2000 1µs to 500S; WS3000 1µs to
1000S}. Parameter is invalid when CARR,WVTP is NOISE or GATE_NCYC
is GATE. Parameter is valid only when TRSR is INT.
STPS
<start phase>
Where <start phase>:= {0 to 360}. Parameter is invalid when CARR,WVTP is
NOISE or PULSE.
GATE_NCYC
{GATE|NCYC}
Configures burst waveform either gated or N-cycled. Parameter is invalid
when CARR,WVTP is NOISE.
TRSR
{EXT, INT, MAN}
Sets trigger source EXTernal, INTernal, or MANual. Parameter is valid only
when GATE_NCYC is NCYC.
DLAY
<delay>
Where <delay>:= {default unit is S; min. 0; max. depends on the Burst
period}. Parameter is valid only when GATE_NCYC is NCYC.
PLRT
{NEG|POS}
Sets polarity NEGative or POSitive. Parameter is valid only when
GATE_NCYC is GATE.
TRMD
<trigger mode>
Where <trigger mode>:= {RISE, FALL, OFF}. Parameter is invalid when
TRSR is EXT. Parameter is valid only when GATE_NCYC is NCYC.
EDGE
{RISE|FALL}
Use rising or falling edge. Parameter is invalid when CARR,WVTP is NOISE.
Parameter is valid only when GATE_NCYC is NCYC.
TIME
<cycle time>
Where <cycle time>:= {1 to 50000}. Parameter is valid only when
GATE_NCYC is NCYC.
MTRIG
Triggers the device once. Parameter is valid only when TRSR is MAN.
CARR,WVTP
<wave type>
Where carrier <wave type>:= {SINE, SQUARE, RAMP, PULSE, NOISE,
ARB}. CARR,WVTP cannot be NOISE if GATE_NCYC is NCYC.
(CARR,)FRQ
<frequency>
Where carrier wave <frequency>:= {default unit is Hz; minimum 1xe-6 Hz,
maximum depends on the WaveStation model}.
922829 Rev C
17