REV 3 EZCT-2000 USER’S MANUAL

53

5.2

Verifying the EZCT-2000’s V

x

Sense Circuit Using an External Meter

The excitation voltage (V

x

) sensed by the EZCT-2000 can be verified using an external RMS volt

meter. Follow the steps below to verify the EZCT-2000’s V

x

sense circuit:

a. Connect the X cables to an RMS volt meter as shown in Figure 15.

b. Select the EZCT-2000’s Diagnostics mode (see section 5.1).

c. Press the

[PAPER Contrast]

key to raise the V

x

voltage.

d. Verify the V

x

voltage displayed on the EZCT-2000 LCD screen against the value displayed

on the external volt meter.

e. Press the

[STOP]

key to end the test.

Figure 15. EZCT-2000 V

x

Verification Test Connections