5.3

Verifying the EZCT-2000’s I

x

Sense Circuit Using an External Meter

You can verify the excitation current (I

x

) sensed by the EZCT-2000 by using an external RMS

ampere meter. Follow the steps below to verify the EZCT-2000’s I

x

sense circuit:

a. Connect the X cables to a power resistor and an RMS ampere meter as shown in Figure

16.

b. Select the EZCT-2000’s Diagnostics mode (see section 5.1).

c. Press the

[PAPER Contrast]

key to raise the V

x

voltage.

d. Verify the I

x

voltage displayed on the EZCT-2000’s LCD screen against the value

displayed on the external RMS ampere meter.

e. Press the

[STOP]

key to end the test.

Figure 16. EZCT-2000 I

x

Verification Test Connections