Operating controls and indicators, power, charger on – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual

Page 18: batt. ready (autotest), fault, testing, test button, Setup, Battery charging procedure (autotest), 2shudwlqj &rqwurov dqg ,qglfdwruv

Advertising
background image

ZOLL Base PowerCharger

4x4

Operator’s Manual

14

2SHUDWLQJ &RQWUROV DQG ,QGLFDWRUV

The ZOLL Base PowerCharger

4x4

with AutoTest

automatically tests battery capacity with each battery
recharge.
The BATT. READY indicator light illuminates when the
battery is fully charged and capable of powering an
M Series equipped with Pace/Defibrillation and Pulse
Oximetry (SpO

2

) for approximately 1.5 hours in Monitor

mode. Fully charged batteries whose capacity is
insufficient will cause a FAULT light to illuminate.
With AutoTest, the full charging cycle is complete in
eight (8) hours or less for the PD 4410/Smart/
Smart Ready Battery and fifteen (15) hours or less for all
XL Batteries.

 32:(5

Illuminates when the ZOLL Base PowerCharger

4x4

is

connected to live AC mains and the unit is ready for
charging and/or testing.

 &+$5*(5 21

Illuminates when a battery is properly installed in the
corresponding charging/testing compartment and the
charge cycle is in progress.
Illuminates with the TESTING light during AutoTest and
recharge of the battery.
There are four CHARGER ON indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above it.

 %$77 5($'< $XWR7HVW

Illuminates at the end of a test/charge cycle. Indicates
that the battery has been charged to 100% of its present
capacity and is ready for use.
Illuminates at the end of a battery test or battery
compartment test and remains lit as long as the battery
remains in the compartment. Indicates that the battery
has passed the capacity test and has sufficient capacity
to power an M Series equipped with Pace/Defibrillation
and Pulse Oximetry (SpO

2

) for approximately 1.5 hours

in Monitor mode, or that the battery compartment is
operating properly.
There are four BATT. READY indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 )$8/7

Illuminates with the TESTING light when a battery fails
the capacity test. Also, illuminates when a battery
compartment or any other failure occurs.

There are four FAULT indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 7(67,1*

Illuminates when either a battery capacity test cycle is in
progress or a battery compartment test is completed.
Illuminates with the BATT. READY light indicating a fully
charged battery, capable of powering an M Series
equipped with Pace/Defibrillation and Pulse
Oximetry (SpO

2

) for approximately 1.5 hours in Monitor

mode.
Illuminates with the CHARGER ON light during AutoTest
and recharge of the battery.
Illuminates with the FAULT light to indicate a battery that
has failed the capacity test.
There are four TESTING indicator lights. Each one
corresponds to the battery charging/testing compartment
directly above.

 7(67 EXWWRQ

Initiates battery test cycle or battery compartment test.
The battery capacity test cycle automatically charges,
discharges and recharges the battery.
If a battery is installed in the battery compartment and
the corresponding TEST button is pressed, a battery
capacity test is performed. If the battery compartment is
empty and the corresponding TEST button is pressed, a
compartment test is performed and an audible tone will
sound, indicating that the test has passed.
There are four TEST buttons. Each one corresponds to
the battery charging/testing compartment directly above.

6HWXS

Place the ZOLL Base PowerCharger

4x4

on a stable

secure surface. There must be at least four (4) inches of
clear space around the unit to allow for proper heat
dissipation during battery charging or testing. Do not
place anything on top of or beneath the unit, such as
blankets or cloths. Doing so may block the vents on the
unit, preventing proper dissipation of heat during
operation.
Turn the unit on by plugging the line cord into the power
module at the rear of the unit and then into live AC
mains.

%DWWHU\ &KDUJLQJ 3URFHGXUH $XWR7HVW

The ZOLL PD 4410/Smart/Smart Ready Battery may be
charged in any of the four battery compartments.
XL Batteries should only be charged in a ZOLL Base
PowerCharger

4x4

labeled “XL Battery Ready”. Each

battery charge compartment operates independently of

Advertising