5

General Information

Specifications

ZOLL Base PowerCharger

4x4

General

Indicator Lights

Controls

ZOLL Batteries

Size

4.2 “ H x 13.2 “ W x 12.2 “ L inches

Weight

4.1 kg (9.0 lbs) without batteries.

Equipment Type IEC Type B

Power
Requirements

100-120 VAC, 50/60 Hz input.
220-240 VAC, 50/60 Hz input.

Power
Consumption

100 watts maximum.

Design
Standards

Designed to meet or exceed UL 544,
EN 60601, and CSA 22.2 standards for
medical equipment safety.

Recharge Time

QuickCharge

Total recharge of fully depleted battery:

4 hours or less for the PD 4410/Smart/
Smart Ready Battery.

8 hours or less for all XL Batteries.

AutoTest

AutoTest and total recharge of fully
depleted battery:

8 hours or less for the PD 4410/Smart/
Smart Ready Battery

15 hours or less for all XL Batteries

Temperature

5 °C to 40 °C (operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non
condensing).

External Fuses

2 Type T, 2A/250V Time Delay fuses
located above AC mains input.

POWER

Illuminates when ZOLL Base
PowerCharger

4x4

is plugged into live AC

mains.

CHARGER ON

Indicates battery is being charged.
Separate indicator light for each battery
compartment.

BATT.READY

Illuminates when the charge cycle is
complete. Separate indicator light for
each battery compartment.

FAULT

Illuminates when battery capacity test or
battery compartment test fails. Separate
indicator light for each battery
compartment.

TESTING

Illuminates when either a battery test or a
battery compartment test is in progress.
Separate indicator light for each battery
compartment.

ZOLL Base PowerCharger

4x4

with

AutoTest only:
Illuminates with the CHARGER ON light
during AutoTest and recharge of the
battery

TEST Button

Illuminates TESTING indicator light and
initiates battery test cycle or battery
compartment test.
Automatically charges, discharges and
recharges the battery.
Separate indicator light for each battery
compartment.

Type

Rechargeable, sealed lead acid:

ZOLL Battery Pack PD4410
ZOLL Smart Ready Battery
ZOLL Smart Battery
ZOLL XL Smart Ready Battery
ZOLL XL Smart Battery

Weight

PD 4410/Smart/Smart Ready Battery: 1 kg
(2.2 lbs)
XL Battery: 1.68 kg (3.7 lbs)

Size

PD 4410/Smart/Smart Ready Battery:

4 cm H x 7 cm W x 18.9 cm L
(1.6 in. x 2.8 in. x 7.4 in.)

XL Battery:

9.4 cm H x 7 cm W x 20 cm L
(3.7 in. x 2.8 in. x 7.9 in.)

Capacity

PD 4410/Smart/Smart Battery:

2.5 Amp Hour rating for new batteries

XL Battery:

4.5 Amp Hour rating for new batteries

Nominal Battery
Voltage

2.0 V/cell; 5 cells.

Temperature

0 °C to 50 °C (operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non
condensing).

Capacity Test

At a minimum, test batteries every
three (3) months. Refer to ZOLL’s Battery
Management Program brochure.