Battery compartment testing procedure, Dwwhu\ &rpsduwphqw 7hvwlqj 3urfhgxuh – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual

Page 20

Advertising
background image

ZOLL Base PowerCharger

4x4

Operator’s Manual

16

2. Insert a battery into any compartment. Make sure the

battery is fully seated (a click can be heard.)

3. Press the TEST button for the appropriate

compartment. The TESTING indicator light will
illuminate for the duration of the test cycle.

4. If the battery passes the test, the BATT. READY light

will illuminate while the TESTING light remains lit,
indicating that the battery is fully charged, ready for
use, and has adequate capacity to power an
M Series equipped with Pace/Defibrillation and Pulse
Oximetry (SpO

2

) for approximately 1.5 hours in

Monitor mode.

5. If a battery fails the test, the FAULT light will

illuminate while the TESTING light remains lit. The
unit will beep when a battery that has failed capacity
test is removed.
To identify the defect, press and hold the TEST
button while the battery is still in the compartment.
Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 17 and
take corrective action.
Repeat test.
Before disposing of the battery, test the battery
compartment for possible defective operation (see
“Battery Compartment Testing Procedure” on
page 16).

6. Record the date and results of the test on the battery

label.

1RWH Interrupting a test cycle in progress by removing

the battery will cause an audible warning of
five (5) consecutive beeps to indicate that the
battery is not fully charged and should not be
used. All four indicator lights will illuminate for
four (4) seconds and then go off.

%DWWHU\ &RPSDUWPHQW 7HVWLQJ

3URFHGXUH

The ZOLL Base PowerCharger

4x4

has four battery

compartments that are able to test and charge batteries
simultaneously and independently of each other. In
addition, the ZOLL Base PowerCharger

4x4

can test each

battery compartment for defective operation through the
simple press of a button.
To test a battery compartment:
1. Check to see that the unit is connected to live AC

mains. The POWER light should be lit.

2. Remove battery from compartment to be tested.
3. Press TEST button under battery compartment.
4. All four indicator lights will illuminate for

approximately 4-5 seconds.

5. If the TESTING light and BATT. READY light remain

lit for three (3) seconds while the other lights go off,
the compartment is fully operational without any
defects.

6. If the FAULT light is lit, the compartment is defective

in some way.
To identify the defect, press and hold the TEST
button. Illumination of certain indicator lights together
represent a specific defect. Refer to “Fault Codes
and Definitions” on page 17 to id
entify the fault and
take corrective action.

1RWH Battery compartments operate independently of

each other and one faulty compartment may not
affect any others. However, there are common
failures that can affect all compartments.

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

Advertising
This manual is related to the following products: