15

Operation (AutoTest)

the others. Four batteries may be charged
simultaneously. Installation or removal of adjacent
batteries will not affect the ongoing charging of batteries
in other compartments. Correct placement of the battery
into any compartment will automatically initiate the
charging cycle.

To turn the ZOLL Base PowerCharger

4x4

on, plug the

power cord into live AC mains. All indicator lights will
illuminate for approximately five (5) seconds and then go
off. An audible tone will sound indicating self-test passed
and the POWER light will remain lit.

To charge a battery:
1. Check to see that the unit is connected to live AC

mains. The POWER light should be lit.

2. Insert battery into any compartment. Make sure the

battery is fully seated (a click can be heard).

3. The CHARGER ON and TESTING indicator lights

illuminate and remain on during the charging cycle.

4. In eight (8) hours or less for the PD4410/Smart/

Smart Ready Battery and fifteen (15) hours or less for
all XL Batteries, the CHARGER ON light will go off
and the BATT. READY light will illuminate indicating
that the battery has passed the capacity test, is fully
charged and is ready for use. The capacity test
indicates that the battery has adequate capacity to
power an M Series equipped with Pace/Defibrillation
and Pulse Oximetry (SpO

2

) for approximately

1.5 hours in Monitor mode.

If the battery or battery compartment is not capable of
charging, the FAULT light will illuminate and the
CHARGER ON light may stay on.

To identify the defect, press and hold the TEST
button while the battery is still in the compartment.
Illumination of certain indicator lights together
represent a specific defect; see “Fault Codes and
Definitions” on page 17 and
take corrective action.
Before disposing of the battery, test the battery
compartment for defective operation (see “Battery
Compartment Testing Procedure” on page 16) and
perform a battery test.

1RWH Interrupting a charge cycle in progress by

removing the battery will cause an audible
warning of five (5) consecutive beeps to indicate
that the battery is not fully charged and should not
be used. All four indicator lights will illuminate for
four (4) seconds and then go off.
Repeated use of partially charged batteries will
reduce battery capacity and operating life.

5. To remove the battery, insert a finger into the recess

at the top of the battery compartment, press against
the battery to disengage the battery locking clip and
lift the battery out. A compartment test is performed
illuminating all indicator light, briefly followed by the
BATT. READY and TESTING indicator lights
illuminating for two (2) seconds and then all indicator
lights going off, except the POWER indicator light.

0DQXDO %DWWHU\ 7HVWLQJ 3URFHGXUH

Although this test is available, the AutoTest feature
eliminates the need to manually test batteries since
battery capacity is automatically tested with each battery
recharge.
All four battery compartments can test batteries. It is not
necessary to insert a fully charged battery for testing; the
ZOLL Base PowerCharger

4x4

will automatically charge

the battery first, test it, and then recharge the battery to
full capacity. Testing a battery in one compartment will
not affect any other charge or testing cycles in progress.
To test a battery:

1. Check to see that the ZOLL Base PowerCharger

4x4

is connected to live AC mains. The POWER light
should be lit.

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

TEST

TESTING

FAULT

BATT. READY

CHARGER ON

Advertising