Tone parameters, Qrs volume on power-up, Low battery tone – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 20
Advertising

Chapter 2
Configuration Parameters
14
9650-1201-01 Rev. M
Tone Parameters
The Tone parameters control audio functions on the E Series unit, and include:
•
QRS Volume on Power-Up
•
Low Battery Tone
QRS Volume on Power-Up
This parameter allows you to select the default QRS beeper volume following power on of the
unit. If you set this parameter to Off, the unit does not beep during ECG monitoring.
Values are as follows:
Low Battery Tone
This parameter allows you to set the interval between Low Battery warning beeps. The Low
Battery warning message is displayed once per minute, regardless of this setting.
Values are as follows:
Possible Values:
On, Off
Default:
On
Possible Values:
1 minute, 5 minutes
Default:
1 minute