12 lead secs per lead, 12 lead print speed, Ptca message enabled – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 39

12-Lead Parameters
9650-1201-01 Rev. M
33
12 Lead Secs Per Lead
This parameter allows you to set the number of seconds of ECG data printed for each lead on
the 12-lead 4x3 strip. This setting affects only 12-lead reports generated by pressing the
Acquire softkey.
Values are as follows:
12 Lead Print Speed
This parameter allows you to set the print speed of the 12-lead 4x3 reports. This setting affects
only 12-lead reports generated by pressing the Acquire softkey.
Values are as follows:
PTCA Message Enabled
This parameter allows you to review, or make changes to, Percutaneous Transluminal Coronary
Angioplasty (PTCA) settings if the E Series unit includes the 12-Lead Reperfusion Algorithm
option. It determines whether or not the E Series unit prints a PTCA message (if appropriate).
Values are as follows
:
Possible Values:
2.5, 5.0, 7.5, 10.0 seconds
Default:
2.5 seconds
Possible Values:
25 mm/second, 50mm/second
Default:
25 mm/second
Possible Values:
Yes, No
Default:
Yes