Specifications – ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual

Page 21

Advertising
background image

9650-1212-01 Rev. E

19

Specifications

Specifications

This section summarizes the specifications of the E Series End Tidal Carbon Dioxide (EtCO

2

) option.

CAPNOSTAT 5® CO

2

Sensor

LoFlo™

Transducer Type

Mainstream

Sidestream

Principle of Operation

Non-Dispersive Infrared (NDIR) single beam optics, dual wavelength, no moving parts.

Warm Up Time

Full specifications within 2 minutes at an ambient temperature of 25º C. Capnogram in
15 seconds.

EtCO

2

Measurement Range

0 - 150 mmHg
0 - 20%
0 - 20 kPa

EtCO

2

Accuracy

(at 760 mmHg, ambient
temperature of 25°C)

0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
101 - 150 mmHg, ±10% of reading

(At respiration rates > 80 breaths per minute, all ranges are ±12% of actual.)

EtCO

2

Resolution

0.1 mmHg 0 - 69 mmHg
0.25 mmHg 70 - 150 mmHg

EtCO

2

Stability

Short-Term Drift: Drift over four hours < 0.8 mmHg.

Long-Term Drift: Accuracy specification will be maintained over a 120 hour period after
zeroing.

EtCO

2

Noise

RMS noise of the sensor < 0.25 mmHg at
7.5% CO

2

.

RMS noise of the sensor < 0.25 mmHg at
5% CO

2

.

EtCO

2

Rise Time (10-90%)

< 60 ms (Adult/pediatric adapters)

< 60 ms (Infant/pediatric adapters)

< 3 seconds (includes transport and rise
time)

Respiration Rate Range

2 - 150 breaths per minute

Respiration Rate Accuracy

±1 breath per minute

Sample Flow Rate

N/A

50 ml/min ±10 ml/min

Compensations

Barometric pressure 400 - 850 mmHg (automatic).
Operator selectable O

2

/N

2

O compensation.

EtCO

2

Alarm Limits

User selectable, High 5 - 100 mmHg, Low 0 - 95 mmHg, OFF.

Respiration Rate (RR) Alarm
Limits

User selectable, High 5 - 150 respirations per minute, Low 0 - 100 respirations per
minute, OFF.

Halogenated Agents

Specification allows for halogenated anesthetic agents that may be present at normal
clinical levels. The presence of Desflurane in the exhaled breath beyond normal values
(5-6%) may positively bias carbon dioxide values by up to an additional 2-3 mmHg.

Airway Adapter Deadspace

Adult < 5 cc
Infant < 1.0 cc

Adult < 7.3 cc Maximum
Pediatric/Infant < 1.0 cc

Environmental (sensor or
module)

Operating Temperature: 0º C to 45º C (0º C to 40º C for LoFlo Module)
Storage and Shipping Temperature: -40º C to 70º C

The E Series unit may not perform to specifications when stored at the upper or lower
extreme limits of storage temperature and immediately put into use.

Electromagnetic Immunity

AAMI DF-80:2003, IEC 60601-1-2, 10 V/m