ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual

Page 9

Advertising
background image

Sidestream EtCO2 Setup

9650-1212-01 Rev. E

7

Selecting a Sidestream Cannula

Select a sidestream cannula based on the patient's size
and monitoring situation. Nasal and nasal/oral cannulas
are disposable and single patient use.

Inserting the Sample Cell

Follow these steps:

1. Remove the LoFlo sampling cannula or airway

adapter kit from the package.

2. Insert the LoFlo sample cell into the LoFlo sample

cell receptacle and ensure that it clicks into place.

3. Ensure that the LoFlo module exhaust tube vents

gases away from the module environment.

4. Turn the selector switch on the E Series to

MONITOR (ON for AED units).

5. Wait for the CO

2

module to warm up.

The unit will display the WARM UP message for
approximately one minute while the module warms
up to operating temperature. The message
disappears when the module is ready for use.

Note Warm up time varies with ambient temperature of

the module.

Zeroing the LoFlo CO

2

Module/Sample Cell

The module/sample cell zero allows the LoFlo CO

2

module to adjust to the optical characteristics of the
sample cell. While zeroing is recommended the first time
a particular LoFlo module is connected to the unit, it is
only absolutely necessary when the message ZERO
CO2 MODULE
is displayed.

CAUTION! Always ensure that the sample cell is

properly connected to the LoFlo module
before zeroing.

1. Ensure that the nasal cannula or on-airway adapter is

not connected to the patient or close to any source of
CO

2

(including the patient's -- and your own --

exhaled breath and ventilator exhaust valves).

2. Press the Param. softkey and select the EtCO

2

menu

item, then press Enter.

3. Press the Zero softkey.

The unit zeroes the module and displays the
ZEROING CO2 MODULE message for
approximately 15-20 seconds.

The units displays the message ZERO DONE upon
completion of the zeroing.

Note Do not attempt zeroing for 20 seconds after

removing the adapter or cannula from the patient’s
airway. This time allows any CO

2

remaining in the

adapter or cannula to dissipate before zeroing. Do
not attempt to zero the module while the adapter or
cannula is in the patient’s airway. Zeroing with CO

2

in the adapter or cannula can lead to inaccurate
measurements and/or other error conditions. If you
attempt zeroing while CO

2

remains in the adapter

or cannula, the time required to zero the module
may be increased. If zeroing cannot be completed,
the message “ZERO FAILED” will be displayed. If
this occurs, clear any occlusion in the adapter or
cannula, remove the source of CO

2

, wait 20

seconds, and try zeroing again.

Applying a Sidestream Airway Adapter Kit

The sidestream airway adapter kit is intended for
monitoring the EtCO

2

of intubated patients.

Before attaching the airway adapter to the breathing
circuit, verify that the adapter is clean, dry, and
undamaged. Replace if necessary.

Cannula Application

Nasal CO

2

Sampling Cannula, Adult

Nasal CO

2

sampling only

Nasal CO

2

Sampling Cannula,

Pediatric

Nasal CO

2

Sampling Cannula, Infant

Oral/Nasal CO

2

Sampling Cannula,

Adult

Oral/Nasal CO

2

sampling only

Oral/Nasal CO

2

Sampling Cannula,

Pediatric

Nasal CO

2

Sampling with O

2

Delivery

Cannula, Adult

Nasal CO

2

sampling with
oxygen delivery

Nasal CO

2

Sampling with O

2

Delivery

Cannula, Pediatric

Oral/Nasal CO

2

Sampling with O

2

Delivery Cannula, Adult

Oral/Nasal CO

2

sampling with
oxygen delivery

Oral/Nasal CO

2

Sampling with O

2

Delivery Cannula, Pediatric

Module Cable

Sample Cell
Receptacle

LoFlo C5 Module

Exhaust
Tube

Figure 5

LoFlo Module