Safety considerations warnings – ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual

Page 4

Advertising
background image

E Series - End Tidal Carbon Dioxide (EtCO

2

) Option Insert

2

9650-1212-01

Rev.

E

Safety Considerations

WARNINGS

General

Carefully read the E Series Operator’s Guide and
these operating instructions before operating the
EtCO

2

monitoring option.

Ensure that the E Series EtCO

2

option is operated

by qualified personnel only.

Do not use the E Series EtCO

2

option as an apnea

monitor.

Do not immerse the E Series unit, patient cables, or
sensors in water, solvents, or cleaning solutions.

If the accuracy of any reading is suspect, first check
the patient’s vital signs by alternate means and then
check the E Series EtCO

2

option for proper

operation.

If an alarm condition occurs while the alarms are
suspended, the suspended alarm indications will
only be visual displays and symbols. No audio alarm
indications will occur.

Elevated oxygen levels, nitrous oxide, or
halogenated agents contained in the breathing
gases may degrade the accuracy of measurements
made with the E Series EtCO

2

option. Activate

oxygen compensation if O

2

levels in excess of 60%

are introduced. Activate N

2

O compensation if nitrous

oxide is introduced into the airway circuit. The
presence of Desflurane beyond 5% may positively
bias the carbon dioxide reading by up to 3 mmHg.

Do NOT use the LoFlo module on patients who
cannot tolerate the removal of 50ml/min of breathing
gases from the airway.

Carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.

Do not touch the bed, patient, or any equipment
connected to the patient during defibrillation. A
severe shock can result. Do not allow exposed
portions of the patient’s body to come in contact with
metal objects, such as a bed frame, as unwanted
pathways for defibrillation current may result.

CAPNOSTAT and Accessories

Always ensure the integrity of the patient breathing
circuit after insertion of the airway adapter by
verifying a proper CO

2

waveform (capnogram) on

the monitor display.

Do not use CAPNOSTAT 5 or LoFlo sensors in the
presence of flammable anesthetics or other
flammable gases.

Do not attempt to open the sensor. An electrical
shock hazard exists internally. Refer servicing to
qualified personnel.