Specifications (cont.) – ZOLL M Series CCT Defibrillator Rev E User Manual
Page 14
8
ZOLL M Series CCT
Specifications (cont.)
Table: Environmental Specifications
Operating Temperature:
32° to 122° F (0° to 50° C)
Storage and Shipping Temperature:
-4° to 140° F (-20° to 60° C)
NOTE: The M Series CCT device may not perform to specifications
when removed from storage at either the upper or lower
extreme temperature limits and immediately put into use.
Humidity:
5 to 95% relative humidity, noncondensing.
Vibration:
Mil Std 810 E, Minimum Integrity Test
Shock:
IEC 68-2-27, 50g 6mS half sine
Operating Pressure:
594 to 1060 millibars
Material Ingress:
IEC 529, IP23
Electromagnetic Compatibility (EMC):
CISPR 11 Class B - Radiated and Conducted Emissions
Electromagnetic Immunity:
AAMI DF-2: 1996, EN 61000-4-3: 2002, 15 V/m
Electrostatic Discharge:
AAMI DF-2: 1996, EN 61000-4-2: 1995
Conducted Susceptibility:
IEC 61000-4-4: 1995, EN 61000-4-5: 1995, EN 61000-4-6: 1996