ZOLL M Series CCT Defibrillator Rev E User Manual

Page 18

Advertising
background image

12

ZOLL M Series CCT

Recommended Separation Distances from RF Equipment for

the M Series CCT Life-Support Functions

Recommended separation distances between portable and mobile RF communications equipment

and the M Series CCT.

The life-support functions

a

of the M Series CCT are intended for use in the electromagnetic environment in

which radiated RF disturbances are controlled. The customer or user of the M Series CCT can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the M Series CCT as recommended below, according to the
maximum output power of the communications equipment.

a. The life-support functions on the M Series CCT are defined to be any function associated with ECG monitoring,
pacing, defibrillation, and shock analysis. Specifically, these functions include, but are not limited to, the ECG
waveform monitoring from leads or pads, the pacing pulse output, QRS detection, defibrillation energy discharge, and
shock advisory functions.

Rated maximum

output power of

equipment

(in watts)

Separation distance according to frequency of transmitter

(in meters)

150 kHz to 80 MHz

outside ISM bands

150 kHz to 80 MHz

in ISM bands

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.035

0.12

0.06

0.12

0.1

0.11

0.38

0.19

0.38

1

0.35

1.2

0.6

1.2

10

1.1

3.8

1.9

3.8

100

3.5

12

6

12

For transmitters rated at a maximum output power not listed above, the recommended separation distance

d

in

meters can be determined using the equation applicable to the frequency of the transmitter, where

P

is the

maximum output power rating of the transmitter in watts according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
NOTE 3: An additional factor of 10/3 is used in calculating the recommended separation distances for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to
2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if
it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.

d =0.35

P

d =1.2

P

d =0.6

P

d =1.2

P

Advertising