Using the m series cct options, Changing the display traces – ZOLL M Series CCT Defibrillator Rev E User Manual

Page 9

Advertising
background image

ZOLL M Series CCT

3

Using the M Series CCT options

The following sections describe how to use the M Series CCT option.

Changing the Display Traces

Trace 1 always displays the ECG waveform. You may select or change the waveform display for

Traces 2 and 3. If the unit has the appropriate options, you can choose from the following parameters:

• 3 Lead ECG (uses all three display channels)

• IBP Channel 1 (P1)

• IBP Channel 2 (P2)

• EtCO

2

• SpO

2

When the unit is in PACER or DEFIB mode, display messages take the place of Trace 3.

Any changes made remain in effect until either the settings are changed or for 10 seconds after the

M Series is turned off (the 10 second interval allows settings to remain in effect when the battery is

changed).

Trace 2 and 3 are factory configured to be OFF when the unit is powered up. However, they can be

configured to display specific parameters on power-up (refer to the M Series CCT Configuration

Guide).

To select the displayed waveforms:

1.

Press the

Traces

softkey (if the “Traces” softkey label is not displayed, press the

Return

softkey until “Traces” displays, and press the

Traces

softkey).

The Traces menu display is shown in Fig. 3:

Figure 3: Traces Menu

2.

Press the

Select

softkey to select “3 Lead ECG”, “Set Trace 2”, or “Set Trace 3.”

3.

Press the

Enter

softkey.

4.

If “3 Lead ECG” is selected:

• Traces 2 and 3 are set to ECG with each lead based on the active lead group. Trace 1 remains

set with the LEAD hard key. For example, if Trace 1 is set to Lead II with the LEAD hard

key, Trace 2 and 3 to are set to Lead I and Lead III respectively.

• For units with 12 Lead ECG option, Custom Lead groups may be set based on preconfigured

choices. For additional information, refer to the M Series CCT Configuration Guide and 12

Lead Operator’s Guide.