Monitoring (Cont.)

SpO2

1. Connect the SpO2 cable to the patient sensor and to

the blue SpO2 port on the rear of the unit labeled SpO2.

2. Attach sensor to patient’s finger.
3. Turn selector switch to grey area labeled MONITOR on

the R Series ALS. If using BLS or Plus device, turn ON,
then press Manual then Confirm softkeys to use
in ALS mode.

4. If properly attached, the SpO2 values will display

on the screen.

5. To display waveform: Press Options softkey.

Press Traces softkey. Press Trace 2 or Trace 3
softkey. Select SpO2 softkey.

NIBP

1. Plug NIBP cable into the back of the unit in silver port marked NIBP.
2. Attach appropriate BP cuff to patient’s arm or thigh.
3. Turn selector switch to grey area labeled MONITOR. If using BLS or Plus device,

press Manual then Confirm softkeys to use in ALS mode.

4. For ALS: Press the Param softkey, then press the NIBP soft key.
5. On BLS or Plus device, press illuminated blood pressure icon on

bottom left to take measurement.

Alarms

1. Press the Alarms softkey to select alarms.
2. Press Return softkey to activate alarms.
3. To mute for 90 seconds: Press ALARM SUSPEND button.
4. To reactivate muted alarm: Press ALARM SUSPEND button.
5. To disable: Press and hold ALARM SUSPEND for 3 seconds.

Note: Alarms are still active, but silent.