Pediatric defibrillation – ZOLL R Series Monitor Defibrillator Rev E User Manual

Page 7

Advertising
background image

Page 4

Adult Defibrillation – Pediatric Defibrillation

R Series ALS Model

Defibrillation (Advisory)

1. Apply electrodes.
2. Turn selector switch to DEFIB.
3. The energy will be displayed as configured by

your institution.

4. Shout “Clear, Analyzing”; stop CPR and

ensure patient is motionless.

5. Press ANALYZE button.
6. Read display message. If SHOCK ADVISED

is displayed, shout “Clear” and verify others
are clear.

7. Press SHOCK button.
8. If NO SHOCK ADVISED, check for a pulse

and if no pulse, resume CPR immediately.

Defibrillation (Manual)

1. Apply electrodes.
2. Turn selector switch to red area labeled

DEFIB.

3. Energy selected will be as configured by

institution protocol. You may change settings
if desired by using the ENERGY SELECT

▲ (UP) and ▼ (DOWN) arrow keys.

4. If the rhythm appears to be shockable, press

the CHARGE button.

5. Shout “Clear” and verify others are clear.
6. Press SHOCK button.

Pediatric Defibrillation

Follow the instructions as outlined above for your unit with the following exceptions:
• OneStep Pediatric electrodes should be used on children up to 8 years of age. For children over

8 years of age, use adult electrodes.

• When the OneStep Pediatric electrode is connected, the R Series will automatically set

the energy delivered to 50, 70, and 85 J increments. These levels are appropriate for children
from 25 kg or greater.

• When using OneStep Pediatric electrodes on children under 25 kg, an energy of 2 J per kg

is recommended and will need to be set manually using the ENERGY SELECT

▲ (UP) and

▼ (DOWN) arrow keys.

• For all other electrodes, select the desired energy manually based upon 2 J per kg.
• Do not use AED or Advisory mode in children under 8 years of age unless OneStep Pediatric

electrodes are attached or energy has been manually selected in advisory mode.