Adult defibrillation r series bls/plus model – ZOLL R Series Monitor Defibrillator Rev E User Manual

Page 6

Advertising
background image

Page 3

Adult Defibrillation

R Series BLS/Plus Model

Defibrillation (AED)

1. Apply electrodes.
2. Turn selector switch to ON.
3. System will begin to analyze or

instruct PERFORM CPR, depending
upon configuration chosen.

4. Once analysis begins, stand clear.
5. If rhythm is shockable, defibrillator

will charge and red SHOCK button
will illuminate.

6. If shock button illuminates, shout “Clear” –

once all are clear, press SHOCK.

7. If no shock advised, start CPR immediately.

Defibrillation (Advisory)

1. Apply electrodes.
2. Turn selector switch to ON.
3. Press Manual then Confirm softkeys.
4. Press ANALYZE to start advisory function

and follow the instructions.

5. If rhythm is shockable, defibrillator

will charge and red SHOCK button
will illuminate.

6. If shock button illuminates, shout “Clear” –

once all are clear, press SHOCK.

7. If no shock advised, start CPR immediately.

Defibrillation (Manual)

1. Apply electrodes.
2. Turn selector switch to ON.
3. Press Manual, then Confirm softkeys. System will switch to ALS mode

(depending upon configuration, your defibrillator may not require a second key
press to activate ALS mode).

4. Energy select will automatically display the correct adult energy per configuration.
5. If desired, energy may be adjusted up or down.
6. If rhythm appears shockable, press CHARGE button.
7. When red SHOCK button illuminates, shout “Clear” and verify others are clear.
8. Press SHOCK button.
9. Check pulse – if no pulse, resume CPR immediately.