Triggers – Agilent Technologies N9010A User Manual

Page 27

Advertising
background image

Chapter 1

27

Agilent EXA Signal Analyzer

Frequency and Time

Description

Specifications

Supplemental Information

Triggers

Additional information on some of the
triggers and gate sources

Video

Independent of Display Scaling and
Reference Level

Minimum settable level

−170 dBm

Useful range limited by noise

Maximum usable level

Highest allowed mixer level

a

+ 2 dB

(nominal)

Detector and Sweep Type
relationships

Sweep Type = Swept

Detector = Normal, Peak,
Sample or Negative Peak

Triggers on the signal before detection,
which is similar to the displayed signal

Detector = Average

Triggers on the signal before detection, but
with a single-pole filter added to give
similar smoothing to that of the average
detector

Sweep Type = FFT

Triggers on the signal envelope in a
bandwidth wider than the FFT width

RF Burst

Level Range

−50 to −10 dBm plus attenuation
(nominal)

b

Level Accuracy

±2 dB + Absolute Amplitude Accuracy
(nominal)

Bandwidth (

−10 dB)

16 MHz (nominal)

Frequency Limitations

If the start or center frequency is too close
to zero, LO feedthrough can degrade or
prevent triggering. How close is too close
depends on the bandwidth listed above.

Advertising