General – Agilent Technologies N9010A User Manual
Page 67

Chapter 1
67
Agilent EXA Signal Analyzer
General
General
Description
Specifications
Supplemental Information
Calibration Cycle
2 years
Description
Specifications
Supplemental Information
Temperature Range
Operating
a
a. For earlier instruments (S/N prefix <MY/SG/US5052), the operating temperature ranges from 5 to
50
°C.
Altitude
≤ 2,300 m
0 to 55
°C
Altitude = 4,500 m
0 to 47
°C
Derating
b
b. The maximum operating temperature derates linearly from altitude of 4,500 m to 2,300 m.
Storage
c
c. For earlier instruments (S/N prefix <MY/SG/US5052), and installed with hard disk drives, the storage
temperature ranges from –40 to +65
°C.
−40 to +70°C
Altitude
d
d. For earlier instrument (S/N prefix <MY/SG/US5052), the altitude was specified as 3,000 m (approxi-
mately 10,000 feet).
4,500 m (approx 15,000 feet)
Humidity
Relative humidity
Type tested at 95%, +40°C
(non-condensing)
Description
Specifications
Supplemental Information
Environmental and Military
Specifications
Samples of this product have been type tested
in accordance with the Agilent Environmental
Test Manual and verified to be robust against
the environmental stresses of Storage,
Transportation and End-use; those stresses
include but are not limited to temperature,
humidity, shock, vibration, altitude and power
line conditions. Test Methods are aligned
with IEC 60068-2 and levels are similar to
MIL-PRF-28800F Class 3.