General – Agilent Technologies N9010A User Manual

Page 67

Advertising
background image

Chapter 1

67

Agilent EXA Signal Analyzer

General

General

Description

Specifications

Supplemental Information

Calibration Cycle

2 years

Description

Specifications

Supplemental Information

Temperature Range

Operating

a

a. For earlier instruments (S/N prefix <MY/SG/US5052), the operating temperature ranges from 5 to

50

°C.

Altitude

≤ 2,300 m

0 to 55

°C

Altitude = 4,500 m

0 to 47

°C

Derating

b

b. The maximum operating temperature derates linearly from altitude of 4,500 m to 2,300 m.

Storage

c

c. For earlier instruments (S/N prefix <MY/SG/US5052), and installed with hard disk drives, the storage

temperature ranges from –40 to +65

°C.

−40 to +70°C

Altitude

d

d. For earlier instrument (S/N prefix <MY/SG/US5052), the altitude was specified as 3,000 m (approxi-

mately 10,000 feet).

4,500 m (approx 15,000 feet)

Humidity

Relative humidity

Type tested at 95%, +40°C
(non-condensing)

Description

Specifications

Supplemental Information

Environmental and Military
Specifications

Samples of this product have been type tested
in accordance with the Agilent Environmental
Test Manual and verified to be robust against
the environmental stresses of Storage,
Transportation and End-use; those stresses
include but are not limited to temperature,
humidity, shock, vibration, altitude and power
line conditions. Test Methods are aligned
with IEC 60068-2 and levels are similar to
MIL-PRF-28800F Class 3.

Advertising