88

Chapter 3

VXA Vector Signal and WLAN Modulation Analysis Application

Analog Modulation Analysis (N9064A-1FP/1TP)

Analog Modulation Analysis (N9064A-1FP/1TP)

Description

Specs & Nominals

Supplemental Information

AM Demodulation

(Span

≤ 12 MHz,

Carrier

≤ −17 dBFS)

Demodulator Bandwidth

Same as selected measurement span

Modulation Index Accuracy

±1%

Harmonic Distortion

−55 dBc

Relative to 100

% modulation

index

Spurious

−60 dBc

Relative to 100

% modulation

index

Cross Demodulation

0.5

% AM on an FM signal with

50 kHz modulation rate,

200 kHz deviation

PM Demodulation

(Deviation < 180

°,

modulation rate

≤ 500 kHz,

span

≤ 12 MHz)

Demodulator Bandwidth

Same as selected measurement span,

except as noted

Modulation Index Accuracy

±0.5°

Harmonic Distortion

0.5

%

Spurious

−60 dBc

Cross Demodulation

1

° PM on an 80% modulation index

AM signal, modulation rate

≤ 1

MHz