The following table shows possible countermeasures – Yokogawa RotaMASS 3-Series User Manual

Page 65

Advertising
background image

IM 01R04B08-00E-E 1st edition February 15, 2012 -00

All Rights Reserved. Copyright © 2012, Rota Yokogawa

<8. TECHNICAL DETAILS OF MODBUS COMMUNICATION>

8-3

The following table shows possible countermeasures :

Alarm (A);

Warning

(W)

Error (E)

Name

Meaning

Description

Countermeasure

A-01

MF 1 criteria

Mass flow L Alarm

Mass flow below MF 1 criteria

Increase mass flow / Change setting

A-02

MF 2 criteria

Mass flow H Alarm

Mass flow above MF 2 criteria

Reduce mass flow / Change setting

A-03

VF 1 criteria

Volume flow L Alarm

Volume flow below VF 1 criteria

Increase volume flow / Change setting

A-04

VF 2 criteria

Volume flow H Alarm

Volume flow above VF 2 criteria

Reduce volume flow / Change setting

A-05

Dens. 1 criteria Density L Alarm

Density below Dens. 1 criteria

Increase density / Change setting

A-06

Dens. 2 criteria Density H Alarm

Density above Dens. 2 criteria

Reduce density / Change setting

A-07

Temp. 1 criteria Temperature L Alarm

Temperature below Temp. 1 criteria

Increase temperature / Change setting

A-08

Temp. 2 criteria Temperature H Alarm

Temperature above Temp. 2 criteria

Reduce temperature / Change setting

A-09

Conc. 1 criteria Concentration L Alarm

Concentration below Conc. 1 criteria Increase concentration / Change setting

A-10

Conc. 2 criteria Concentration H Alarm

Concentration above Conc. 2 criteria Reduce concentration / Change setting

A-11

Netflow 1 crit.

Net flow L Alarm

Net flow below Net flow 1 crit.

Increase net flow / Change setting

A-12

Netflow 2 crit.

Net flow H Alarm

Net flow above Net flow 2 crit.

Reduce net flow / Change setting

A-13

Total thr exceed Total threshold exceed

Totalizer value exceeds total
threshold

Reset totalizer / Change setting

A-14

Slug detection

Slug flow detected

Slug criterion is reached

Reduce gas bubbles in medium

A-15

Empty pipe
det.

Empty pipe detectd

Empty pipe criterion is reached

Fill pipe

A-16

Corrosion det.

Corrosion detected

Corrosion criterion is reached

Change detector

W-01

Density low

Density lower 0.3 kg/l

Density lower than 0.3 kg/l

Fill the detector with fluid

W-02

Fixed dens act. Fixed Dens selected

Parameter Dens fix val sel is enabled

Set parameter Dens fix val sel to
inhibit

W-03

Fixed temp act. Fixed Temp selected

Parameter Temp fix val sel is enabled

Set parameter Temp fix val sel to
inhibit

W-04

Fixed MF active Fixed MF selected

Parameter MF fix val sel is enabled

Set parameter MF fix val sel to inhibit

W-05

Autozero running Autozero is running …

Autozero process is performed,
started by user action or status input

Wait until process stops
(30sec or 180sec max)

W-06

AZ val error

Autozero value out of
range

Auto zero value is out of the internal
detector depending range

Stop flow during auto zero.
Check detector installation

W-07

AZ fluct error

Autozero fluctuation out
of range

Auto zero fluctuation is out of the
internal detector depending range

Stop flow during auto zero. Check
electrical installation, vibrations and
density.

W-08

Sensor simul.

Sensor simulation active

Fixed frequency or fixed phase dif-
ference is enabled

Set Sensor simulation to inhibit

W-09

Fixed Prim Out Analog output 1 fixed

Analog output 1 fixed at 4mA for
multidrop operation

Set HART polling address to 0
(Multidrop operation stops)

W-10

Prim Out sat.

Analog output 1 saturated

Analog output 1 reaches less than
3.8mA or more than 20.5mA

Increase or decrease current level of
output 1

W-11

Sec Out sat.

Analog output 2 saturated

Analog output 2 reaches less than
3.8mA or more than 20.5mA

Increase or decrease current level of
output 2

W-12

Tert Out sat.

Pulse/Status output 1
saturated

Relation between pulse width and
pulse rate of Pulse/Status output 1
is too high

Decrease length of pulse width or
increase amount of pulse rate of pulse
output 1

W-13

Quart Out sat.

Pulse/Status output 2
saturated

Relation between pulse width and
pulse rate of Pulse/Status output 2
is too high

Decrease length of pulse width or
increase amount of pulse rate of pulse
output 2

E-01

Frequency fault Frequency fault detected

Driving frequency out of range

Check cable (Remote) and detector

E-02

Signal fault

Signal fault detected

Phase difference out of range

Check cable (Remote) and detector

E-03

EEPROM fault EEPROM fault detected

CRC-Error

Contact Yokogawa service

E-04

CPU fault

CPU fault detected

Micro processor is not working

Contact Yokogawa service

E-05

DSP fault

DSP fault detected

Micro processor is not working

Contact Yokogawa service

E-06

Sensor 1 fault

Sensor 1 Signal fault
detected

Sensor 1 signal below 7% for 3 min Check cable (Remote) and detector

E-07

Sensor 2 fault

Sensor 2 Signal fault
detected

Sensor 2 signal below 7% for 3 min Check cable (Remote) and detector

E-08

Temp sens fault

Temperature Sensor Fault
detected

T<-220°C or T>550°C

Check cable (Remote) and detector
Check temperature