<8. Maintenance>

8-9

IM 01C27H01-01EN

Connect the device configuration tool and check self-diagnostics.

Contact Yokogawa service personnel.

F0808.ai

Refer to Alarm Message Summary

in Subsection 8.5.3

Bind capillaries together as far as

possible or provide lagging.

Provide lagging and/or heat insulation,

or allow adequate ventilation.

Refer to 8.2 “Calibration Instruments

Selection”.

Adjust the output.

Does

the ambient

temperature significantly

differ between the high and low

pressure side

capillaries?

Is output adjusted correctly?

Large output error.

Does the self-

diagnostic indicate the problem

location?

Is transmitter

installed where there is

marked variation in

temperature?

Are excess

capillaries secured?

YES

NO

YES

NO

NO

YES

NO

YES

NO

YES

Secure them so that they are not

moved by wind or vibration.

Were appropriate

instruments used for

calibration?

YES

NO