<10. General Specifications>

10-2

IM 01C27H01-01E

Ambient Humidity Limits

0 to 100% RH

Working Pressure Limits

See table 10.1.

For atmospheric pressure or below, see figure

10.1 and 10.2.

2.7 (0.38)

100 (14.5)

0.1(0.014)

1(0.14)

10 (1.4)

-50 0

50 100 150 200 250 300 350

F01E.ai

Process temperature

for fill fluid code B

Process temperature

for fill fluid code A

Process temperature

for fill fluid code C

Flange max.

working

pressure

Atmospheric

pressure

Transmitter ambient

temperature range

(for fill fluid code

A, B)

Process Temperature (°C)

Working

pressure

kPa abs

(psi abs)

Figure 10.1 Working Pressure and Process

Temperature (Fill fluid : silicone oil for

general and high temperature use)

2.7 (0.38)

0.1(0.014)

1 (0.14)

10(1.4)

100 (14.5)

51 (7.4)

-50 0

50 100 150 200 250 300 350

F02E.ai

Process temperature

for fill fluid code E

Process temperature

for fill fluid code D

Flange max.

working

pressure

Atmospheric

pressure

Transmitter ambient

temperature range

(for fill fluid code D)

Process Temperature (°C)

Working

pressure

kPa abs

(psi abs)

Figure 10.2 Working Pressure and Process

Temperature (Fill fluid : fluorinated

oil for oil-prohibited use and ethylene

glycol for low temperature use)

Table 10.1

Process temperature, Ambient temperature, and Working pressure

T06E.ai

B

C

D

E

–50 to 100°C

(–58 to 212°F)

A

–30 to 180°C

(–22 to 356°F)

10 to 310°C

(50 to 590°F)

–20 to 120°C

(–4 to 248°F)

–10 to 60°C

(14 to 140°F)

–15 to 60°C

(5 to 140°F)

–10 to 250°C

*4

(14 to 482°F)

10 to 60°C

(50 to 140°F)

–40 to 60°C

(–40 to 140°F)

–10 to 60°C

(14 to 140°F)

Specific gravity

*3

1.09

0.94

1.90 to 1.92

1.90

Process temperature

*1

Code

Ambient temperature

*2

2.7 kPa abs

(0.38 psi abs) to

flange rating pressure

1.07

Working pressure

Fluorinated oil

(oil-prohibited use)

Ethylene glycol

(low temperature use)

Silicone oil (general use)

Silicone oil

(high temperature use)

Silicone oil (general use)

51 kPa abs

(7.4 psi abs) to

flange rating pressure

100 kPa abs

(atmospheric pressure)

to flange pressure rating

*1:

See figure 10-1 and 10-2 ‘Working Pressure and Process Temperature.

*2:

This ambient temperature is the transmitter ambient temperature.

*3:

Approximate values at a temperature of 25°C (77°F)

*4:

In case of wetted parts material code TW (Tantalum), process temperature limit is up to 200°C (392°F).

Note: The differential pressure transmitter should be installed at least 600 mm below the high pressure (HP) process connection.

However, this value (600 mm) may be affected by ambient temperature, operating pressure, fill fluid or material of the wetted

diaphragm. Contact YOKOGAWA when the transmitter can not be installed at least 600 mm below the HP process connection..