Figure 4.1, Table 4.1, Installation location – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 8

Advertising
background image

8

IM 11M12A01-01E-A

Figure 4.1

4.2 Installation of the Detector (Model ZR22G-015-Q)

Note: Ceramic (Zirconia) is used in the sensor (cell) portion of the detector probe. Care should be
taken to not drop the detector during installation. The same applies to the High Temperature Adapter
transport tube.
A gasket should be used on the flange surface to prevent gas leakage. The gasket material should be
selected depending on the characteristics of the measured process sample gas. It should be heat and
corrosion resistant to the sample temperatures and gases present. These parts are to be supplied by
the user, and are described in Table 4.1.

Mounting Flange Specification

Accessory Name

Quantity

Note

ZR22G-015 Probe

JIS 5K32A Flange

*Gasket

1

Heat and Corrosion Resistant

Bolt (M12 x 50)

4

Nut (M12)

4

Washer (M12)

8

ZR22P Adapter to Process

ANSI 4.0 inch 150# FF Flange

*Gasket

1

Heat and Corrosion Resistant

Bolt (M16x 60)

8

Nut(M16)

8

Washer (M16)

16

*Specifications for gasket material(s) vary, and are dependent upon process application. A local pipefitting
distributor can assist with proper selection of gasket materials

Table 4.1

4.2.1

Installation Location

The ZR22G-015-Q model detector is used with the ZR22P High Temperature Probe Adapter, when
sample gas temperatures exceed 700ºC (1292ºF). Additionally, this detector can be utilized where
maintenance space is limited.