4 dio module assembly test, Caution – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 20

Advertising
background image

2-11

SM 04L31A01-01E

Testing

2

2.4

DIO Module Assembly Test

This section describes the test procedure for CTRL MODULE ASSY (B8700CY) which is

used on the CX1000/CX2000. This test is not necessary if you perform the general tests

in section 2.2. Perform this test when testing the module by itself.

Test Instruments

Instrument

Specifications

Withstanding voltage tester

AC 1 to 3 kV, 500 VDC

Jig

A measurement instrument having the same functions and
characteristics as the CX1000/CX2000

Tests

• Withstand Voltage Test

Consult your nearest Yokogawa representative regarding the following tests.

• DI Test

• DIO Test

Testing Environment

Ambient temperature of 23

±

5

°

C, relative humidity of 55

±

20%

CAUTION

• The following assumes that the test will be performed with the module installed in

the CX1000/CX2000.

• The test cannot be performed on models having no internal control loops with no

DIO module installed.

Test Procedure

Perform a withstand voltage test on the module installed in the CX1000/CX2000 at the

points listed below.

Test Points

Reference Values

DO terminal to CX1000/CX2000 earth terminal*

Leakage current of 2 mA or less at 0.5 kV AC for 1 minute

DI terminal to CX1000/CX2000 earth terminal

Leakage current of 2 mA or less at 0.5 kV AC for 1 minute

*

Short DO1—12 and C.

† Short all DIs.

Advertising