Chapter 3 Adjustments

3.1

Calibration of the Measuring Instrument’s Input

Allow the instrument to warm up for at least 30 minutes prior to calibration.

Instruments for Calibration

DC voltage generator

Accuracy: 0.005% of setting +1

µ

V

Resistor

100

, 300

(accuracy: 0.01% or less)

Overview

The structure of the A/D converter in the CX2000 differs depending on the model.

Multiple channels (groups) share one A/D converter. All A/D converters must be

calibrated, so calibration must be performed on each group.

CX1000

Monitor

No. of Inputs for

Measurement

Total Number

No. of

No. of

Group

Meas.

Interval

of Channels

CH/AD

A/D

6ch

1s

6

6

1

(1—6)

*

Group: a classification for all channels that share a single A/D converter

Control (1st Slot)

Number of Internal

Measurement

Total Number

No. of

No. of

Group

Control Loops

interval

of Channels

CH/AD

A/D

0 loops

0

0

0

2 loops

250 mS

5

5

1

(1—5)

CX2000

Control (1st Slot)

Number of Internal

Measurement

Total Number

No. of

No. of

Group*

Control Loops

interval

of Channels

CH/AD

A/D

0 loops

0

0

0

2 loops

250 mS

5

5

1

(1—5)

4 or 6 loops

250 mS

10

5

2

(1—5) (6—10)

* Group: a classification for all channels that share a single A/D converter

Monitor (2nd, 3rd Slot)

No. of Inputs for

Measurement

Total Number

No. of

No. of

Group

Meas.

Interval

of Channels

CH/AD

A/D

10 ch

1s

10

10

1

(1—10)

20 ch

1s

20

10

2

(1—10) (11—20)

Ranges to Be Calibrated and Their Parameters

Range

Zero

Full Scale

20 mV

0 mV

20 mV

60 mV

0 mV

60 mV

200 mV

0 mV

200 mV

1 V

0 V

1V

2 V

0 V

2 V

6 V

0 V

6V

20 V

0 V

20 V

Pt100

100

300

† You must calibrate the 200 mV range before calibrating the RTD range.