2 control output calibration, Control output calibration -6 – Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 26

Advertising
background image

3-6

SM 04L31A01-01E

3.2

Control Output Calibration

Allow the instrument to warm up for at least 30 minutes prior to calibration.

Instruments for Calibration

DC current meter

Accuracy: 0.01% of rdg

Resistor

250

±

1%

Overview of Calibration

During calibration of the control output (current output terminal), each output is

connected to one D/A converter, so you must calibrate the outputs separately. The

module installed in the CX1000 is 01. In the CX2000, the modules are numbered in

order (01, 02, 03) from top to bottom. Also within each module, output number 01 is

assigned to loop 1, and output 02 to loop 2.

Calibration Procedure

The calibration procedure can be carried out using keys or communications commands.

Calibration Using Keys

1.

Connect a 250

Ω ±

1% resistor and DC current meter in serial to the current

output terminal.

2.

Turn ON the unit while holding down the UP arrow key. The calibration screen

appears.

3.

Press the Pid Analog Output Adjust soft key. The Pid Analog Output Adjust

screen is displayed.

4.

Use the arrow keys and soft keys to select the module number and output

number to be calibrated.

5.

While monitoring with a current meter, change the calibration value at each Output

Value (4 mA and 20 mA) so that it falls within the calibration range shown in the

following chart. The first digit of the calibration value corresponds to 0.366211

µ

A.

Output Value

Calibration Range

4 mA

3.99 mA—4.01 mA

20mA

19.99 mA—20.01 mA