IM 04L65B01-01EN

4.9 Things to Consider

4.9.1 Time Zone and Daylight Saving Time

Be sure to set the same time zone and daylight saving

time settings on the PC running the GA10 server, the PC

running the GA10 client, and the data collection devices.

If they are not the same, data time may not be displayed

correctly.

4.9.2 Error Data

If collected or recorded data is in error, it is displayed

or recorded using indications other than values. For the

different types of error data, see “Data that indicates

errors.”

• Data Display in a Digital Monitor Set or Meter

Monitor Set

Display

Data Condition

*

+OVER

+OVER

–OVER

-OVER

INVALID

INVALID

BURNOUT

BURNOUT

ILLEGAL

ILLEGAL

LACK

LACK

OFF

OFF

* See “Data that indicates errors.”

• Display in the Trend Monitor Set

Waveform

Cursor Value Data Condition

*

Drawn exceeding the

scale upper limit

+OVER

+OVER

Drawn exceeding the

scale lower limit

-OVER

-OVER

Nothing

INVALID

INVALID

BURNOUT

BURNOUT

ILLEGAL

ILLEGAL

LACK

LACK

(blank)

OFF

* See “Data that indicates errors.”

• Data in Recording Data Files

Data in Binary Data

Files

Data in Excel Data

Files

Data Condition

*

+OVER

+OVER

+OVER

–OVER

–OVER

-OVER

INVALID

INVALID

INVALID

BURNOUT

BURNOUT

BURNOUT

ILLEGAL

ILLEGAL

ILLEGAL

LACK

LACK

LACK

OFF

OFF

OFF

* See “Data that indicates errors.”

• Data That Indicates Errors

The following table shows the different types of data

that indicates errors.

Data

Description

+OVER

+Over-range data

–OVER

-Over-range data

SKIP

Channels that have been set to skipped

INVALID

Invalid data

The data type and decimal place specified on

the Tag Setting Page do not match those of the

collected data.

BURNOUT Burnout data

ILLEGAL

Illegal data

LACK

Indicates that the device failed to acquire the data

OFF

Indicates one of the following conditions.

• Data collection has not been performed since

the project was opened.

• Channels are not assigned to tags.

• When the data time is set to PC time, the

collected data is SKIP data.

• Communication error condition

• Initialized condition as a result of changing the

Device Setting Page or Tag Setting Page while

data collection is stopped

• An attempt was made to collect data from a

device using the backfill function, but there is

no data recorded in the device.

4.9.3 Reflecting Changes Made on the Monitor

Page to the Setting Page

If the access privilege is Owner or Manager, changes

made to the following settings on the Monitor Page are

reflected on the corresponding Setting Page (Display

Group or Acquisition & Monitor). If the access privilege is

Operator or Monitor, the changes are not reflected.

• Monitor Set size adjustment

• Waveform display on/off state, Y-axis display on/

off state, Detail/Compact, Zoom in/Zoom out, and

movement in the Trend Monitor Set

• Trip line position

4.9.4 Changing the Time on the Device after

Starting Data Collection and Recording

Do not change the time on the device after starting data

collection and recording, because doing so will cause

adverse effects on the monitor screen and recorded data.

Related topic “Changes to devices during data

collection and recording”:

Q11