70 •

Channels

Within each wireless group, the devices communicate
with each other via three channels (C1, C2, C3). This
ensures reliable, troublefree data transmission.

When you set up a wireless group with this scale, the
device will suggest three channels guaranteeing opti-
mum data transmission. We recommend accepting the
channel numbers suggested.

You can also select the channel numbers (0 to 99)
manually - for example if you want to set up more than
one wireless group.

The channels must be sufficiently far apart to ensure
trouble-free data transmission. We recommend a
spacing of at least 30 between channel numbers. Each
channel number may only be used for one channel.

Example configuration: channel numbers when setting
up 3 wireless groups within one surgery:

• wireless group 0: C1=_0, C2=30, C3=60

• wireless group 1: C1=10, C2=40, C3=70

• wireless group 2: C1=20, C2=50, C3=80

Detecting devices

If you wish to set up a wireless group with the scale, it
will search for other active devices from the

seca 360°

wireless

system. The devices detected are shown as

modules in the display on the scale (e.g. “MO 3”). The
meaning of the numbers is as follows:

• 1: personal scale

• 2: length measuring rod

• 3: wireless printer

• 4: PC with seca USB wireless adapter

• 7: baby scale

• 5, 6 and 8-12: reserved for system expansion