Maintenance/recalibration, 1 information on maintenance and recalibration – Seca 757 User Manual

Page 79

Advertising
background image

Maintenance/recalibration • 79

English

11.MAINTENANCE/RECALIBRATION

11.1 Information on maintenance and recalibration

We recommend having your device serviced prior to
recalibration.

ATTENTION!

Incorrect measurements as a result of poor
servicing

– Have servicing and repairs carried out

exclusively by seca service or by an
authorized service partner.

– You can find service partners in your area at

www.seca.com or by sending an e-mail to
[email protected].

Have an authorized technician perform recalibration
according to national legal regulations. The year of first
calibration can be found after the CE symbol on the rat-
ing plate above the number of the notified body 0109
(Hessische Eichdirektion [Hessen Weights & Measures
Office]).

Recalibration is necessary whenever one or more cali-
bration seals are damaged or the contents of the cali-
bration counter no longer match the number on the
applicable calibration counter sticker. If calibration
seals are damaged, contact seca service directly.

… the display “Er:X:16”
appears?

Oscillations have been induced in the scale, pre-
venting determination of the zero point.

- Do not touch the trolley or table supporting

the scale.

- Restart scale.

… when the Enter key
(send/print) is pressed,
the “Er:x:71” display
appears?

Data transmission not possible, wireless module is
disabled.

- Activate wireless module (“Activating the wire-

less module (SYS)” on page 71).

… when the Enter key
(send/print) is pressed,
the “Er:x:72” display
appears?

Data transmission is not possible, no wireless
group set up.

- Set up wireless group (“Setting up a wireless

group (Lrn)” on page 71).

Fault

Cause/Remedy

Advertising