2 disinfecting, 3 sterilisation, Function check – Seca 757 User Manual

Page 76

Advertising
background image

76 •

8.2

Disinfecting

ATTENTION!

Damage to device

The viewing windows of scales and displays are
made of acrylic glass. Acrylic glass windows
can crack or become cloudy if unsuitable disin-
fectants are used on them.
– Only use disinfectants suitable for sensitive

surfaces. Suitable disinfectants are available
from specialist dealers.

1. Follow the instructions on the disinfectant.

2. Disinfect the device at regular intervals using a soft

cloth dampened with a suitable disinfectant.

3. Please take note of the following terms:

8.3

Sterilisation

This device may not be sterilised.

9. FUNCTION CHECK



Perform a function check prior to each use.

A complete function check includes:

• visual inspection for mechanical damage

• checking the alignment of the device

• visual and function check of the display elements

• function check of all the controls shown in the

section entitled “Overview”

• function check of optional accessories

If you notice any faults or deviations during the function
check, first try to resolve the error with the aid of the
section entitled “What do I do if ...” in this document.

Interval

Component

Before every measurement

Tray

After every measurement

Tray

If required

Display elements,
controls, housing