What do i do if – Seca 757 User Manual

Page 77

Advertising
background image

What do I do if …? • 77

English

CAUTION!

Personal injury

If you notice any faults or deviations during the
function check which cannot be resolved with
the aid of the section entitled “What do I do if
...” in this document, you may not use the
device.
– Have the device repaired by seca service or

by an authorized service partner.

– Follow the section entitled “Servicing/

recalibration” in this document.

10.WHAT DO I DO IF …?

Fault

Cause/Remedy

… no weight is dis-
played when scale is
loaded?

The scale is not connected to the power supply.

- Check if the scale is switched on.
- Check whether the battery block is inserted

and that the plug contacts are properly con-
nected.

… 0.000 does not appear
before weighing?

A load was placed on the scale before it was
switched on.

- Remove load.
- Switch scale off and back on again.

… a segment lights up
continuously or not at
all?

The relevant segment is faulty.

- Please notify maintenance service.

is

displayed?

Battery block power is running down.

- Charge battery block.

… the display “bAtt”
appears?

Battery block discharged.

- Charge battery block.

… the display “StOP”
appears?

The maximum load has been exceeded.

- Remove load.

… the display “tEMP”
appears?

The ambient temperature of the scale is too high
or too low.

- Set up scale in an ambient temperature

between +10 °C and +40 °C.

- Wait approx. 15 minutes until scale has

adjusted to ambient temperature.