Cleaning, What to do if – Seca 959 User Manual

Page 83

Advertising
background image

Cleaning • 83

English

8. CLEANING

Clean the seat and frame on an as needed basis using
a household cleaner or a commercial disinfectant. Fol-
low the manufacturer's instructions.

Do not use scouring or caustic cleaning products,
spirit, petrol or similar substances for cleaning under
any circumstances. Such products may damage the
high-quality finishes.

9. WHAT TO DO IF …?

Malfunction

Cause/Remedy

… no weight is displayed
during weighing?

The scale has no power supply.

- Check whether the scale is switched on
- Check whether the rechargeable battery

block is inserted

0.00

does not appear

before weighing?

A load was placed on the scales before being
switched on.

- Remove load
- Switch scales off and back on again

… a segment lights up
continuously or not at
all?

There is a fault at that point.

- Notify maintenance service.

is displayed?

The voltage of the rechargeable battery block
drops.

- Charge the rechargeable battery block as

soon as possible

bAtt

is displayed?

Rechargeable battery block is empty.

- Charge rechargeable battery block

St0p

is displayed?

The maximum load has been exceeded.

- Reduce load

is displayed?

The ambient temperature of the scales is too high
or too low.

- Set up scales in an ambient temperature

between +10 °C and +40 °C

- Wait approx. 15 minutes until scales have

adjusted to ambient temperature

Advertising
This manual is related to the following products: