Technical data • 87

English

Medical device according to directive
93/42/EEC

Class I with measuring function

Calibration according to directive 2009/23/EC Class III

EN 60601-1:

• Class II totally insulated appliance:

• Type B electromedical device:

Maximum load

• Partial weighing range 1
• Partial weighing range 2
• Partial weighing range 3

150 kg
250 kg
300 kg

Minimum load

• Partial weighing range 1
• Partial weighing range 2
• Partial weighing range 3

1 kg
2.0 kg
4.0 kg

Graduations

• Partial weighing range 1
• Partial weighing range 2
• Partial weighing range 3

50 g
100 g
200 g

Tare range

300 kg

Accuracy following initial calibration

• Partial weighing range 1, 0 kg up to 25 kg
• Partial weighing range 1, 25 kg up to

100 kg

• Partial weighing range 1, 100 kg up to

150 kg

• Partial weighing range 2, 0 kg up to 50 kg
• Partial weighing range 2, 50 kg up to

200 kg

• Partial weighing range 2, 200 kg up to

250 kg

• Partial weighing range 3, 0 kg up to 100 kg
• Partial weighing range 3, 100 kg up to

300 kg

±25 g
±50 g

±75 g

±50 g
±100 g

±150 g

±100 g
±200 g

Radio transmission

• Frequency band
• Transmission power
• Standards applied

2.433 GHz - 2.480 GHz
< 10 mW
EN 300 328
EN 301489-1
EN 301489-17