Van Air Systems Dew Point Meter 46-2529 2-02 User Manual

Dew point meter

Advertising
background image

INSTALLATION, OPERATION, & MAINTENANCE MANUAL

FOR

DEW POINT METER

P/N 46-2529

WARNING

BEFORE STARTING INSTALLATION AND/OR MAINTENANCE PROCEDURES, TURN OFF POWER AND DEPRESSURIZE
UNIT BEFORE SERVICING TO PREVENT SERIOUS INJURY. SERIOUS PERSONAL INJURY MAY RESULT IF THESE SAFETY
RULES ARE NOT FOLLOWED.

DO NOT REMOVE, REPAIR, OR REPLACE ANY ITEM ON THIS UNIT WHILE IT IS ENERGIZED OR UNDER PRESSURE. TURN
OFF MAIN POWER TO THE UNIT AND DEPRESSURIZE THE SAMPLE CELL BY CLOSING THE ISOLATION VALVE
COMPLETELY BEFORE STARTING ANY INSTALLATION OR MAINTENANCE PROCEDURES.

WHEN MAKING ANY ELECTRICAL CONNECTIONS TO THE METER, COMPLY WITH ALL LOCAL, STATE, AND FEDERAL
CODES I.E. THE NATIONAL ELECTRICAL CODE. ALWAYS USE COMPONENTS AND WIRING RATED FOR THIS
INSTALLATION.

INTRODUCTION

SECTION 1

1.1 SPECIFICATIONS

Dew Point Range ........ -40

o

F to +15

o

F (-40

o

C to -9

o

C)

Accuracy ..................... +2

o

F

Alarm Set Point ........... -10

o

F (-23

o

C) (Adjustable)

Remote Alarm Output

Normally open and closed contacts
5A @ 115V rated

Recorder Output ......... 4-20 mA or 0-5V (jumper selectable)

Power Requirement ..... 115VAC (+10%) 50/60Hz

Power Consumption ... 4 Watts

Enclosure .................... Nema 4, Fiberglass w/window

1.2 DIMENSIONS

INSTALLATION

SECTION 2

2.1 INSTALLATION

The meter can be mounted on any flat vertical surface. The
enclosure has threaded inserts on the back for mounting.

Once the unit has been mounted, the sample line can be connected
to the sample cell isolation needle valve. The sample line should be
1/4" copper tubing. The pressure of the sample line should not
exceed 150 PSIG.

The unit will require 115 volts for operation. Run the necessary
conduit and wiring to the enclosure.

If the alarm relays and the recorder output are to be used, they will
require several procedures for installation. Otherwise, no additional
installation steps are required to use the meter.

NOTE:
See Model 6392N and 6392N2 instructions for operation,
maintenance, printed circuit board details, and calibration of
equipment.