3.

Connect power to terminals HOT+ and NEUT- and ground to terminal GND of TB1 as shown on
drawing.

4.

Connect Sensor Cable to terminals A and B of TB1.

5.

Recorder output signal is on TB3. Terminal 1 of TB3 is 0-5V output , Terminal 3 is

4-20mA output and Terminal 2 is signal ground.


Wiring Summary

Terminal No.TB1 Labeled Function

1

HOT+

Line Hot (or positive if 24V version board)

2

NEUT-

Line Neutral (or negative if 24V version board)

3

GND

Power and signal ground

4

NC1

Normally closed for relay #1

5

COM1

Common for relay #1

6

NO1

Normally open for relay #1

7

NC2

Normally closed for relay #2 (applies only to 6392N2)

8

COM2

Common for relay #2 (applies only to 6392N2)

9

NO2

Normally open for relay #2 (applies only to 6392N2)

10 A

Sensor

lead

11 B

Sensor

lead

12

GND

Power and signal ground

13 C

(not

used)

14 D

(not

used)

Terminal No. TB2

Function

1

Aux alarm +12VDC source

2

Aux alarm open collector output relay #1

3

Aux alarm +12VDC source (applies only to 6392N2)

4

Aux alarm open collector output relay #2 (applies only to 6392N2)

Terminal No. TB3

Function

1

Recorder output 0-5VDC

2

Power and signal ground

3

Recorder output 4-20mA

4

Aux no alarm +12VDC source

5

Aux no alarm open collector output relay #1

6

Aux no alarm +12VDC source (applies only to 6392N2)

7

Aux no alarm open collector output relay #2 (applies only to 6392N2)