Run test script (manual), Ge 91 – GW Instek PSW-Series User Manual User Manual
Page 91

OPERATION
91
Note
If the USB drive is not recognized, check to see
that the function settings for F-20 = 1 (page 106). If
not, reinsert the USB flash drive.
3. Configure T-02 (Test Load) to 1~10
(save memory slot)
T-02 range
1~10 (t001 ~t010)
4. The script will now be available in the memory
slot the script was saved to.
Note
Error messages: If you load a file that is not
present on the USB drive “Err 002” will be
displayed on the display.
Run Test Script (Manual)
Overview
A test script can be run from one of ten memory
slots.
Steps
1. Before a test script can be run, it
must first be loaded into one of the
10 memory save slots.
2. Configure T-01 (Run Test) to 1~10
(save memory slot#)
T-01 range
1~10
3. The test script will automatically start to run.