Configure]:cotime – GW Instek PEL-3000 Series Programming Manual User Manual

Page 49

Advertising
background image

COMMNAND OVERVIEW

47

Parameter/
Return parameter

OFF or 0

Turns the Count Time
timer off.

ON or 1

Turns the Count Time
timer on

Example

[:CONFigure]:CNTime ON
Turns the Count Time timer on.

Query example

[:CONFigure]:CNTime?
>ON
Count Time is turned on.

[:CONFigure]:COTime

Set

Query

Description

Sets or queries the load cutoff time. A cutoff time
of 0 seconds is the equivalent of disabling the
cutoff time.

Syntax

[:CONFigure]:COTime {<NRf>[s] | MINimum |
MAXimum | OFF }

Query Syntax

[:CONFigure]:COTime?

Parameter

<NRf>[s]

Cut off time in seconds
(1~3599999)

OFF

Turns the cutoff time off.

MINIMUM or MIN

Sets the cutoff time to the
maximum

MAXIMUM or MAX

Sets the cutoff time to the
minimum

Return parameter <NR1>

Returns the cutoff time

Example

:COTime MAX
Sets the cutoff time to the maximum.

Query example

:COTime?
>3599999
The cutoff time is set to3599999 seconds.